AC2T research GmbH - Austrian Excellence Center for Tribology
Wiener Neustadt | Website
Short Description
device for the elemental analysis of solid and fluid samples in the range of µg/kg to 100 g/kg
Contact Person
Ing. Rainer Franz
Research Services
realisation of R&D services in the framework of customer orders or long-term cooperation in (inter-)national research projects
Methods & Expertise for Research Infrastructure
elemental analysis for determination of additives, wear and contaminations; a prepared sample solution is introduced as an aerosol into a plasma at a temperature of 6000 to 10000 °C where it is atomised and excited; the excited atoms and ions emit radiation which is detected qualitatively and quantitatively; if necessary, prior to this procedure the sample is submitted to a microwave digestion to avoid matrix effects during the measurement; all chemical elements except halogens, inert gases, hydrogen, oxygen, nitrogen and carbon; qualitative and quantitative analysis